Raw results

From Quantum kot
Jump to navigation Jump to search

21 October 2016

26 October 2016

28 October 2016

Response time of AOM testing. For more information.


31 October 2016

RBW is 5 MHz, VBW 5 MHz

RBW, [MHz] VBW, [MHz] MBR side pwr, [mW] Toptica side pwr, [mW] Pulse duration, [us] Pulse period, [us]
5 5     5 5
31 of October. Pulsed 4WM with the same pumps duration 5 microseconds and period 5 microseconds. This is one shot picture. Sweep time 50us.
31 of October. Pulsed 4WM with the same pumps duration 5 microseconds and period 5 microseconds. This is averaged over many trials. Sweep time 100us.
31 of October. Pulsed 4WM with the same pumps duration 5 microseconds and period 5 microseconds. This is one shot picture. Sweep time 50us.

 

 

1st of November 2016

Measurement conditions.

RBW, [MHz] VBW, [MHz] MBR side pwr, [mW] Toptica side pwr, [mW] Pulse duration, [us] Pulse period, [us]
5 5 0.8 13 5 5

 

We first measured 4WM signal dependence on a seed (Toptica pump entering the fiber). We see the logarithmic increase until saturation, which happens at 10-20 uW. 

1st of November. DLCZ period 20 microseconds, delay between pumps is 150 ns.

4th November 2016

We measured in a pulsed regime dependence of 4WM signal on a side pump MBR. RBW = VBW = 5MHz. Sweep time 50 us. Pulses were 5us long and spaced with 5us. The maximum signal we achieved was ~7mV. Problems:

  • When we look at the signal after locking the cavity it is maximum and then decreases.
  • As well sometimes there is a modulation on top of the signals.
  • Cavity lock is unstable in the pulsed regime: breaks the lock after 3 minutes.