Green Pump

From Quantum kot
Jump to navigation Jump to search

Verdi-V10

S/N: V10-G6017

MFD: 20120131

Verdi's controller

Model: 0178-281-50

Laser control

Coherent GUI

there exists a , Coherent GUI - AVIA, Verdi, Vitesse and Monaco [1]

How to run program

to operate open the following folder and run interpretor:

root@arina-desktop:~/Verdi/verdi-py/src# python Python 2.7.6 (default, Jun 22 2015, 17:58:13) [GCC 4.8.2] on linux2 Type "help", "copyright", "credits" or "license" for more information.

import classes and commands

>>> import Verdi

initialize laser, choose correct comport

>>> laser=Verdi.verdiLaser("/dev/ttyS0")

run commands

>>> laser.getVanadateServo() 'Locked' >>>

Switch on/off procedure

laser.setPower(1) laser.setShutter(0)

Step # Frequency Meaning
1 ? switch stand by -> on position
2 laser.setShutter(1) Shutter open
3 laser.getVanadateServo() Wait till everything is locked (Laser and Vanadate)
4 laser.setPower(5) Set power level 2

Chillers

COM21 -> Verdi Chiller, set temperature is 20 C

COM28 -> MBR chiller, set temperature 18 C


Sample code in interpretator regime:


Useful functions:

setModeStandby()
setModeRun()

Lifetime

Product support stated that expected lifetime is 10.000 (10K) hours, usually they work for 20-25K, once they heard about 100K. Lab C laser (20th of June 2016)

Head [hrs] Diode 1 [hrs] Diode 2 [hrs]
5721.67 3541 3541

Lab A Verdi

Head [hrs] Diode 1 [hrs] Diode 2 [hrs]
8025 11475 11467

Problems

The Verdi controller switches off by itself, we are still try to figure out why. The associated problems with it: The LBO crystal needs to be maintained at ~147. C, as soon as controller if off the crystal temperature decreases to room temperature. So you should proceed with a cold start and wait until the main screen switches from "System warming up" to "Standby" (20-30 min). Regularily if you wish to switch off the oump laser fully, you will use complete shut-down procedure, where you run LBO cool-down cycle.The company designed specifically slow ...

Date Time Comments
4th July 2016 13.09 UPS made a sound, in the wall there was power, controller started to work after we reset the UPS