Green Pump
Verdi-V10
S/N: V10-G6017
MFD: 20120131
Verdi's controller
Model: 0178-281-50
Contents
Laser control
Coherent GUI
there exists a , Coherent GUI - AVIA, Verdi, Vitesse and Monaco [1]
How to run program
to operate open the following folder and run interpretor:
root@arina-desktop:~/Verdi/verdi-py/src# python Python 2.7.6 (default, Jun 22 2015, 17:58:13) [GCC 4.8.2] on linux2 Type "help", "copyright", "credits" or "license" for more information.
import classes and commands
>>> import Verdi
initialize laser, choose correct comport
>>> laser=Verdi.verdiLaser("/dev/ttyS0")
run commands
>>> laser.getVanadateServo() 'Locked' >>>
Switch on/off procedure
laser.setPower(1) laser.setShutter(0)
| Step # | Frequency | Meaning |
|---|---|---|
| 1 | ? | switch stand by -> on position |
| 2 | laser.setShutter(1) | Shutter open |
| 3 | laser.getVanadateServo() | Wait till everything is locked (Laser and Vanadate) |
| 4 | laser.setPower(5) | Set power level 2 |
Chillers
COM21 -> Verdi Chiller, set temperature is 20 C
COM28 -> MBR chiller, set temperature 18 C
Sample code in interpretator regime:
Useful functions:
setModeStandby() setModeRun()
Lifetime
Product support stated that expected lifetime is 10.000 (10K) hours, usually they work for 20-25K, once they heard about 100K. Lab C laser (20th of June 2016)
| Head [hrs] | Diode 1 [hrs] | Diode 2 [hrs] |
|---|---|---|
| 5721.67 | 3541 | 3541 |
Lab A Verdi
| Head [hrs] | Diode 1 [hrs] | Diode 2 [hrs] |
|---|---|---|
| 8025 | 11475 | 11467 |
Problems
The Verdi controller switches off by itself, we are still try to figure out why. The associated problems with it: The LBO crystal needs to be maintained at ~147. C, as soon as controller if off the crystal temperature decreases to room temperature. So you should proceed with a cold start and wait until the main screen switches from "System warming up" to "Standby" (20-30 min). Regularily if you wish to switch off the oump laser fully, you will use complete shut-down procedure, where you run LBO cool-down cycle.The company designed specifically slow ...
| Date | Time | Comments |
|---|---|---|
| 4th July 2016 | 13.09 | UPS made a sound, in the wall there was power, controller started to work after we reset the UPS |